Toggle Main Menu Toggle Search

Open Access padlockePrints

Validation of an electronic image reader training programme for interpretation of [18F]flutemetamol β-amyloid PET brain images

Lookup NU author(s): Professor David BrooksORCiD

Downloads

Full text for this publication is not currently held within this repository. Alternative links are provided below where available.


Abstract

© 2017 The Author(s). Published by Wolters Kluwer Health, Inc. Objectives An electronic training programme (ETP) was developed for interpretation of images during routine clinical use of the PET amyloid imaging agent [ 18 F]flutemetamol injection (VIZAMYL). This study was carried out to validate the ETP. Materials and methods Five nuclear medicine technologists (NMTs) and five readers previously inexperienced in amyloid image interpretation were required to self-train using the ETP and pass a test to participate. A total of 305 [ 18 F]flutemetamol PET images were then tested as the validation set, following preassessment and reorientation (where required) by one of five NMTs. Next, a new set of readers blinded to clinical information independently assessed all 305 images. Images had been acquired in previous studies from patients representing the full spectrum of cognitive capacity. When available, a standard of truth determined by histopathology or clinical history was used to derive sensitivity and specificity for image interpretation from this validation set. Randomly selected images (n=29) were read in duplicate to measure intrareader reproducibility. Images were read first without, and subsequently with anatomic images, if available. Results All NMTs and all readers scored 100% on the qualifying test. The interpretation of 135 cases without anatomic image support resulted in sensitivity ranging from 84% to 94% (majority 94%, median 92%) and specificity ranging from 77% to 96% (majority 92%, median 81%). Inter-reader agreement was very high, with most κ scores more than 0.8. Intrareader reproducibility ranged from 93 to 100%. Conclusion The self-guided ETP effectively trained new amyloid PET image readers to accurately and reproducibly interpret [ 18 F]flutemetamol PET images.


Publication metadata

Author(s): Buckley CJ, Sherwin PF, Smith APL, Wolber J, Weick SM, Brooks DJ

Publication type: Article

Publication status: Published

Journal: Nuclear Medicine Communications

Year: 2017

Volume: 38

Issue: 3

Pages: 234-241

Print publication date: 01/03/2017

Acceptance date: 22/11/2016

ISSN (print): 0143-3636

ISSN (electronic): 1473-5628

Publisher: Lippincott Williams and Wilkins

URL: https://doi.org/10.1097/MNM.0000000000000633

DOI: 10.1097/MNM.0000000000000633


Altmetrics

Altmetrics provided by Altmetric


Share