Toggle Main Menu Toggle Search

Open Access padlockePrints

The use of porous polyethylene (Medpor) lower eyelid spacers in lid heightening and stabilisation

Lookup NU author(s): Christopher Neoh, Dr Alison Dickinson

Downloads

Full text for this publication is not currently held within this repository. Alternative links are provided below where available.


Abstract

Background/aims: The management of lower eyelid retraction can be challenging, and established techniques to correct it are not always successful. Previous reports have suggested a role for the ultrathin high density porous polyethylene lower eyelid spacer (Medpor LES) in such patients. The authors report the experience of three surgeons implanting Medpor LES over 1 year, and ascertain whether such implants are a safe and effective alternative to autogenous spacers. Methods: A prospective, interventional, non-comparative case series of consecutive patients. Surgical indications for Medpor LES were noted. Preoperative and postoperative lower marginal reflex distance (L-MRD), vertical palpebral aperture (PA), lagophthalmos, and scleral show inferior to the limbus (LSS) were recorded, together with major and minor complications. Results: 32 patients (35 eyelids) had a Medpor LES inserted, 22/32 under local anaesthetic, and nine with adjunctive procedures. Mean follow up was 22 months (range 15-28 months). The Medpor LES was effective in reducing the palpebral aperture (p<0.001) and lagophthalmos (p = 0.04) and raising the lower eyelid height by reducing both L-MRD (p = 0.006) and LSS (p<0.001). However there were major complications in 7/32 patients and minor complications in 8/32, most requiring further surgery. Final outcome was good in 24/35 eyelids and satisfactory in 5/35. Conclusions: Despite a good or satisfactory final outcome in the majority of patients, the value of this technique is limited by complications, and should be reserved for those unsuitable for safer techniques.


Publication metadata

Author(s): Tan J, Olver J, Wright M, Maini R, Neoh C, Dickinson AJ

Publication type: Article

Publication status: Published

Journal: British Journal of Ophthalmology

Year: 2004

Volume: 88

Issue: 9

Pages: 1197-1200

ISSN (print): 0007-1161

ISSN (electronic): 1468-2079

Publisher: BMJ Publishing Group

URL: http://dx.doi.org/10.1136/bjo.2003.029397

DOI: 10.1136/bjo.2003.029397


Altmetrics

Altmetrics provided by Altmetric


Share