Browse by author
Lookup NU author(s): Ian Pinder
Full text for this publication is not currently held within this repository. Alternative links are provided below where available.
There are possible advantages of using uncemented fixation in total knee replacement. In this prospective randomised multi-centre study, a comparison was made between cemented and two types of uncemented fixation for the Kinemax design. There were 12-14 cases in each group. Beads were inserted in the bones from which component migration was measured at lime intervals up to 2 years. The axial migrations were significantly less for cemented and HA-coating, compared with press-fit, at all time intervals. The clinical data showed no differences at 2 years except for more cases of pain in the uncemented groups. Radiographically, the cemented interfaces showed the least change, press-fit showed a radiolucent line and a radiodense line, and HA showed a diffuse radiodensity adjacent to the components. It was concluded that for the Kinemax design of tibial component, press-fit was inferior to cemented, but that there was the potential for designing a special component for uncemented fixation for which HA-coating would be an advantage. (C) 2000 Elsevier Science B.V. All rights reserved.
Author(s): Walker PS, Sathasivam S, Cobb A, Learmonth ID, Grobler GP, Pinder IM, Marchetti N, Spinelli MD, Welsby A
Publication type: Article
Publication status: Published
Journal: Knee
Year: 2000
Volume: 7
Issue: 2
Pages: 71-78
ISSN (print): 0968-0160
ISSN (electronic): 1873-5800
Publisher: Elsevier
URL: http://dx.doi.org/10.1016/S0968-0160(99)00041-1
DOI: 10.1016/S0968-0160(99)00041-1
Altmetrics provided by Altmetric