Browse by author
Lookup NU author(s): Dr Colin Kotre, Dr Keith Faulkner, David Rawlings
Full text for this publication is not currently held within this repository. Alternative links are provided below where available.
The authors have constructed a 2D motor-controlled test object phantom holder to simulate clinical situations in which patient movement could be a cause of image degradation. The PAtient MOvement SImulation Test Object (PAMOSITO) has been constructed with modular parts to use different mobile test objects and static structures. The system allows the programming of different cycles of movement along two axes. PAMOSITO has been used in X ray equipment dedicated to interventional radiology. Those systems usually allow for different values for frame rate, pulse width or weighted frame averaging methods. The influence of selecting different values of the parameters, patient movement and its relation to patient dose and image quality has been studied. Image blurring due to motion has been evaluated with Leeds test objects TO.10 and 18FG. Spatial resolution limits and the threshold contrast detail detectability performance have been studied.
Author(s): Kotre CJ; Faulkner K; Rawlings DJ; Guibelalde E; Vano E; Fernandez JM; Ten JI
Publication type: Conference Proceedings (inc. Abstract)
Publication status: Published
Conference Name: 13th International Conference on Solid State Dosimetry (SSD2001)
Year of Conference: 2001
Pages: 155-159
ISSN: 0144-8420
Publisher: Nuclear Technology Publishing
Library holdings: Search Newcastle University Library for this item
Series Title: Radiation Protection Dosimetry
ISBN: