The effect of varying irrigation flow-rate during irrigated radiofrequency ablation on optimising lesion shape

Lookup NU author(s): Dr Rachael Redgrave, Dr Anna Walaszczyk, Micheylla Dewi, Dr Maria Camacho EncinaORCiD, Professor Gavin RichardsonORCiD, Dr Moloy Das

Downloads


Licence

This work is licensed under a Creative Commons Attribution 4.0 International License (CC BY 4.0).


Abstract

Background: Irrigated radiofrequency lesions show a "teardrop" shape, due to endocardial sparing, requiring extensive lesion overlap for contiguity at the endocardial surface. The effect of dynamic irrigation rate on lesion-shape has not been studied.Objective: To assess if dynamic irrigation flowrate during radiofrequency ablation optimises lesion shape by minimising endocardial sparing. Methods:Ex vivo porcine heart model (Phase 1), irrigated ablation lesions were performed in ventricular tissue at 30W with 5 different irrigation protocols A) fixed-rate; B) continuous-reduction; C) continuous-increase; D) stepwise-reduction; E) stepwise-increase. Contact force (10g) and ablation duration (30sec) were constant. Lesion dimensions and endocardial sparing were measured. The optimal protocol was further tested at 20-40W. In Phase 2, the optimal protocol was tested using a in vivo large animal model.Results: Phase 1: Across power settings of 20-40W, continuous reduction in flowrate (B) resulted in decreased endocardial sparing area with maintained lesion depth and volume, compared to fixed-rate irrigation (P<0.001). Phase 2: 65, 30sec lesions were completed in 4 anaesthetised pigs using protocol B (n=25) or fixed-rate irrigation (n=40). The maximum recorded tip temperature was higher for continuous reduction in both the atrium and ventricle but there were no steam pops for either protocol.Conclusions: Compared to fixed-rate, continuous-reduction in irrigation flowrate from 30 to 2ml/min during irrigated radiofrequency ablation results in an optimised lesion shape with reduced endocardial sparing across power settings (20-40W) and proved feasible in an in vivo setting. This may allow for greater lesion spacing while maintaining endocardial contiguity clinically.


Publication metadata

Author(s): Redgrave R, Walaszczyk A, Dewi MW, Encina MC, Clemens J, Matrins R, Richardson GD, Das M

Publication type: Article

Publication status: Published

Journal: EP Europace

Year: 2024

Volume: 26

Issue: 1

Print publication date: 01/01/2024

Online publication date: 27/10/2023

Acceptance date: 11/10/2023

Date deposited: 26/07/2023

ISSN (print): 1099-5129

ISSN (electronic): 1532-2092

Publisher: Oxford University Press

URL: https://doi.org/10.1093/europace/euad321

DOI: 10.1093/europace/euad321

Data Access Statement: The data underlying this article will be shared on reasonable request to the corresponding author.


Altmetrics

Altmetrics provided by Altmetric


Funding

Funder referenceFunder name
Investigator Sponsored Research Grant from Boston Scientific (Ref: ISRRM110379).

Share