Browse by author
Lookup NU author(s): Dr Matthew GermanORCiD
Full text for this publication is not currently held within this repository. Alternative links are provided below where available.
The practical cost considerations of the repeated use and the degree of polishing of the QCM as a commercial viscometer are reported. We have shown that the QCM can be used numerous times with gentle cleaning when used with starch-based foodstuffs with some simple considerations. The change in resonant frequency, Delta(f), and therefore the viscosity response, is not shown to be affected even after 60 applications and cleans. The fundamental unloaded frequency however, showed a linear decrease and therefore the QCM exhibits a mild drop in sensitivity. The degree of polishing of the crystal is shown to be not critical in improving the accuracy of the viscosity response. Although polishing a roughened surface dramatically improves the viscosity response, further polishing does not significantly increase the accuracy. This is shown to be particularly true for low viscosity fluids. The findings here indicate that a viscometer based on the QCM viscometer is a viable alternative to many conventional viscometer methods. (c) 2005 Elsevier Ltd. All rights reserved.
Author(s): Dewar RJ, Ash DC, German MJ, Joyce MJ
Publication type: Article
Publication status: Published
Journal: Journal of Food Engineering
Year: 2006
Volume: 75
Issue: 4
Pages: 461-468
ISSN (print): 0260-8774
ISSN (electronic): 1873-5770
Publisher: Elsevier Ltd
URL: http://dx.doi.org/10.1016/j.jfoodeng.2005.04.032
DOI: 10.1016/j.jfoodeng.2005.04.032