Hip simulator testing of the taper-trunnion junction and bearing surfaces of contemporary modular metal-on-cross-linked polyethylene hip prostheses

Lookup NU author(s): Rohan Bhalekar, Dr Simon Smith, Professor Tom Joyce

Downloads


Licence

This is the final published version of an article that has been published in its final definitive form by John Wiley & Sons, Inc., 2020.

For re-use rights please refer to the publisher's terms and conditions.


Abstract

Adverse reaction to metal debris released from the taper-trunnion junction of modular metal-on-polyethylene (MoP) total hip replacements (THRs) is an issue of contemporary concern. Therefore, a hip simulator was used to investigate material loss, if any, at both the articulating and taper-trunnion surfaces of five 32mm metal-on-cross-linked-polyethylene (MoXLPE) THRs for 5 million cycles (Mc) with a sixth joint serving as a dynamically loaded soak control. Commercially available cobalt-chromium-molybdenum (CoCrMo) femoral heads articulating against XLPE acetabular liners were mounted on 12/14 titanium (Ti6Al4V) trunnions. Weight loss (mg) was measured gravimetrically and converted into volume loss (mm3) for heads, liners and trunnions at regular intervals. Additionally, post-test volumetric wear measurements of the femoral tapers were obtained using a coordinate measuring machine (CMM). The surface roughness (Sa) of femoral tapers was measured post-test. After 5Mc, the mean volumetric wear rate for XLPE liners was 2.74±0.74 mm3/Mc. The CMM measurements confirmed material loss from the femoral taper with the mean volumetric wear rate of 0.045±0.024 mm3/Mc. The Sa on the worn area of the femoral taper showed a significant increase (p<0.001) compared to the unworn area. No other long term hip simulator tests have investigated wear from the taper-trunnion junction of contemporary MoP THRs.


Publication metadata

Author(s): Bhalekar RM, Smith SL, Joyce TJ

Publication type: Article

Publication status: Published

Journal: Journal of Biomedical Materials Research Part B: Applied Biomaterials

Year: 2020

Volume: 108

Issue: 1

Pages: 156-166

Print publication date: 01/01/2020

Online publication date: 29/03/2019

Acceptance date: 11/03/2019

Date deposited: 07/05/2019

ISSN (print): 1552-4973

ISSN (electronic): 1552-4981

Publisher: John Wiley & Sons, Inc.

URL: https://doi.org/10.1002/jbm.b.34374

DOI: 10.1002/jbm.b.34374


Altmetrics

Altmetrics provided by Altmetric


Share