This work aimed to engineer a multi-station shoulder simulator in order to wear test shoulder prostheses using recognized shoulder activities of daily living. A bespoke simulator was designed, built and subject to commissioning trials before a first wear test was conducted. Five JRI Orthopaedics Reverse Shoulder VAIOS 42 mm prostheses were tested for 2.0 million cycles and a mean wear rate and standard deviation of 14.2 ± 2.1 mm3/106 cycles measured for the polymeric glenoid components. This result when adjusted for prostheses diameters and test conditions showed excellent agreement with results from hip simulator studies of similar materials in a lubricant of bovine serum. The Newcastle Shoulder Simulator is the first multi-station shoulder simulator capable of applying physiological motion and loading for typical activities of daily living.
Author(s): Smith SL, Li L, Joyce T
Publication type: Article
Publication status: Published
Journal: Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
Year: 2016
Volume: 230
Issue: 5
Pages: 470-480
Print publication date: 01/05/2016
Acceptance date: 15/09/2015
Date deposited: 24/05/2016
ISSN (print): 0954-4119
ISSN (electronic): 2041-3033
Publisher: Sage Publications Ltd.
URL: http://dx.doi.org/10.1177/0954411915611161
DOI: 10.1177/0954411915611161
Altmetrics provided by Altmetric