In 3D, ultrasound systems sample a conical volume of space with a given angular resolution, resulting in sparsely sampled regions far from the transducer Image analysis operators require that each voxel of a cartesian volume has a value, and some form of interpolation is necessary.3D images of the left ventricle (LV) of 10 patients were constructed from sequences of conventional B-mode frames which were successively rotated about the beam axis. Three different methods of interpolation were used to assign values to unsampled voxels. Reconstructed LV short axis views were extracted from the 3D images and analysed using standard edge detection and threshold operators. Results from the image analysis were used to compare the three different methods of interpolation.Results of this preliminary study show that the choice of interpolation method influences the results of image analysis operations.
Author(s): Sims AJ, Wild J, Pemberton J, Irvine T, Kenny A, Murray A
Publication type: Conference Proceedings (inc. Abstract)
Publication status: Published
Conference Name: Computers in Cardiology 2003
Year of Conference: 2003
Pages: 185-188
Online publication date: 04/05/2004
ISSN: 0276-6547
Publisher: IEEE
URL: https://doi.org/10.1109/CIC.2003.1291121
DOI: 10.1109/CIC.2003.1291121
Library holdings: Search Newcastle University Library for this item
ISBN: 078038170X