Due to their recent introduction there are few studies of retrieved resurfacing hip prostheses. Nine such components associated with groin pain in patients, and five associated with early fracture of the femur, were obtained and analysed using a roundness measuring machine. While the 'fracture' components showed no more than 3 mu m out of roundness, components associated with groin pain showed between 15 and 92 mu m out of roundness values. These latter results indicate wear and correlated with high metal ion levels in these patients, therefore the groin pain was likely associated with an adverse reaction to excessive metal wear debris. (C) 2010 Elsevier Ltd. All rights reserved.
Author(s): Joyce TJ, Langton DJ, Nargol AVF
Publication type: Article
Publication status: Published
Journal: Tribology International
Year: 2011
Volume: 44
Issue: 5
Pages: 517-522
Print publication date: 20/04/2011
Date deposited: 11/01/2012
ISSN (print): 0301-679X
ISSN (electronic): 1879-2464
Publisher: Pergamon
URL: http://dx.doi.org/10.1016/j.triboint.2010.04.006
DOI: 10.1016/j.triboint.2010.04.006
Altmetrics provided by Altmetric